PSN-L Email List Message

Subject: seismometer sensitivity
From: Randall Peters PETERS_RD@..........
Date: Thu, 19 Jul 2012 16:45:29 -0400


There is a great deal to be learned about seismic instrument performance fr=
om consideration of a simple gravitational pendulum.  Even though it was th=
e first instrument of considerable importance, it should not be considered =
an insignificant relic.  Many of you are no doubt familiar with the express=
ion for the period of such a pendulum, oscillating at small amplitudes; i.e=
.., 2 pi times the square root of L/g where L is the length and g is the acc=
eleration of gravity (earth's field, 9.8 m/s^2).   We use this to understan=
d sensitivity by means of the gedanken experiment considered by Einstein in=
 his consideration of relativity.  When the support of the pendulum is acce=
lerated at a constant rate a, the pendulum's angular displacement (radian u=
nits) is simply (a/g).  This means the bob at a distance L from the support=
 is displaced from equilibrium by amount (a L/g).  In turn the length is ex=
pressible as   g  T^2/4 pi^2.   By recognizing that there is always a minim=
um displacement d that can be observed with whatever sensor is used, we sim=
ply obtain the result that the smallest acceleration that can be measured i=
s given by
a =3D  d  ( 2 pi / T)^2  =3D [(2 pi  f )^2 ] d =3D (omega^2) d
This shows that the sensitivity is proportional to the period of the instru=
ment squared.  No matter the nature of the seismograph, this result remains=
 true; i.e., the sensitivity goes as the square of the characteristic perio=
d of the instrument, which ideally is a harmonic oscillator damped with a q=
uality factor of 0.707.
    It is true that the sensor used to detect displacement of the inertial =
mass can be 'tailored' electronically to provide 'useful' frequency depende=
nt characteristics-such as a fairly broad flat response to velocity over a =
range centered on the characteristic frequency.   But no matter what electr=
onics is employed, the reason for deconvolving both (i) the mechanical resp=
onse, and (ii) the response of the electronics employed-is to obtain someth=
ing that is independent of the instrument used.  Only then can meaningful c=
omparisons among different instruments make sense for their observation of =
the same earth motions.
     The following is imperative to understand:  the acceleration response =
of the simple pendulum, or any other seismometer-insofar as the mechanical =
transfer function is concerned-is always flat to ground acceleration below =
the characteristic frequency and falls off as 1/f^2 above that frequency.  =
For electronincs that is flat (independent of frequency), the 'velocity' se=
nsor is one that takes the derivative of bob displacement (as in the coil/m=
agnet Faraday law detector).  This means that above the characteristic freq=
uency, the mechanical fall-off to acceleration is proportional to 1/f, sinc=
e taking the derivative is tantamount to multiplying the otherwise 1/f^2 by=
 f.  In turn a 1/f acceleration response is equivalent to a flat velocity r=
esponse.
(refer to http://www.geophys.uni-stuttgart.de/oldwww/seismometry/man_html/n=
ode12.html)
Consequently, in this range (and this range only)  the mechanical part inde=
pendent of electronics yields an output that is proportional to ground velo=
city.  For this same type instrument, in the frequency range below the char=
acteristic, the mechanical response is proportional to the derivative of ac=
celeration, which is called a 'jerk' .
     I am surprised the extent to which seismologists seem unconcerned with=
 the PSD except when it comes to demonstrating instrument performance under=
 hopefully quiet (somewhat universal background) noise conditions.  Calcula=
ting the PSD for records containing earthquakes I have found to be very inf=
ormative, and I don't know of others having done so.  If I have just not be=
en looking in the right places; please, somebody let me know where to find =
such records.
     The same is also true of the seismocardiography research that I'm invo=
lved with, supported by an NIH grant.  Especially useful in diagnosing hear=
t abnormalities is what one obtains by integrating over the PSD to get the =
cumulative spectral power.   It allows multiple plots to be overlaid on the=
 same graph without the clutter that results when trying to do the same wit=
h PSD's.   If you're interested in this work, just type 'cardiac signals' i=
nto Google w/o the tick marks and click on the 2nd item of the first page.
     Randall


There is a great= deal to be learned about seismic instrument performance from consideration= of a simple gravitational pendulum.  Even though it was the first ins= trument of considerable importance, it should not be considered an insignif= icant relic.  Many of you are no doubt familiar with the expression fo= r the period of such a pendulum, oscillating at small amplitudes; i.e., 2 p= i times the square root of L/g where L is the length and g is the accelerat= ion of gravity (earth’s field, 9.8 m/s^2).   We use this to= understand sensitivity by means of the gedanken experiment considered by E= instein in his consideration of relativity.  When the support of the p= endulum is accelerated at a constant rate a, the pendulum’s angular d= isplacement (radian units) is simply (a/g).  This means the bob at a d= istance L from the support is displaced from equilibrium by amount (a L/g).=   In turn the length is expressible as   g  T^2/4 pi^2.=    By recognizing that there is always a minimum displacement d t= hat can be observed with whatever sensor is used, we simply obtain the resu= lt that the smallest acceleration that can be measured is given by

a =3D  d  ( 2 pi / T)^2  =3D [(2= pi  f )^2 ] d =3D (omega^2) d

This= shows that the sensitivity is proportional to the period of the instrument= squared.  No matter the nature of the seismograph, this result remain= s true; i.e., the sensitivity goes as the square of the characteristic peri= od of the instrument, which ideally is a harmonic oscillator damped with a = quality factor of 0.707.

  &nb= sp; It is true that the sensor used to detect displacement of the inertial = mass can be ‘tailored’ electronically to provide ‘useful&= #8217; frequency dependent characteristics—such as a fairly broad fla= t response to velocity over a range centered on the characteristic frequenc= y.   But no matter what electronics is employed, the reason for d= econvolving both (i) the mechanical response, and (ii) the response of the = electronics employed—is to obtain something that is independent of th= e instrument used.  Only then can meaningful comparisons among differe= nt instruments make sense for their observation of the same earth motions. =

     The follo= wing is imperative to understand:  the acceleration response of the si= mple pendulum, or any other seismometer—insofar as the mechanical tra= nsfer function is concerned—is always flat to ground acceleration bel= ow the characteristic frequency and falls off as 1/f^2 above that frequency= ..  For electronincs that is flat (independent of frequency), the ̵= 6;velocity’ sensor is one that takes the derivative of bob displaceme= nt (as in the coil/magnet Faraday law detector).  This means that abov= e the characteristic frequency, the mechanical fall-off to acceleration is = proportional to 1/f, since taking the derivative is tantamount to multiplyi= ng the otherwise 1/f^2 by f.  In turn a 1/f acceleration response is e= quivalent to a flat velocity response.  

(refer to http://www.geophys.uni-stuttgart.de/oldwww/seismometry/man_= html/node12.html)

Consequently, in this= range (and this range only)  the mechanical part independent of elect= ronics yields an output that is proportional to ground velocity.  For = this same type instrument, in the frequency range below the characteristic,= the mechanical response is proportional to the derivative of acceleration,= which is called a ‘jerk’ . 

     I am surprised the extent to which sei= smologists seem unconcerned with the PSD except when it comes to demonstrat= ing instrument performance under hopefully quiet (somewhat universal backgr= ound) noise conditions.  Calculating the PSD for records containing ea= rthquakes I have found to be very informative, and I don’t know of ot= hers having done so.  If I have just not been looking in the right pla= ces; please, somebody let me know where to find such records. 

     The same is also= true of the seismocardiography research that I’m involved with, supp= orted by an NIH grant.  Especially useful in diagnosing heart abnormal= ities is what one obtains by integrating over the PSD to get the cumulative= spectral power.   It allows multiple plots to be overlaid on the= same graph without the clutter that results when trying to do the same wit= h PSD’s.   If you’re interested in this work, just ty= pe ‘cardiac signals’ into Google w/o the tick marks and click o= n the 2nd item of the first page.

     Randall

 

 

<= /html>=

[ Top ] [ Back ] [ Home Page ]