Hi Geoff.

Increasing the sample rate to 125sps (-3dB at 36Hz), didint solve the issue
and increased the >10s noise..i will try to increase more the sample rate.
But increasing the sample rate reveals a very high count at 50Hz, i check
the earth connections it looks all ok..if i remove the geophone from the
ADC, i have +-2 counts of noise at 16bits and the 50Hz spike is gone.
Simply touching a wire from earth in any part of the geophone, will make
the geophone stiff, without responding.

SAC file with 15minutes at 125SPS
http://dl.dropbox.com/u/15802004/aa.sac

Netherless the 50Hz issue, getting only 3800ohm from a 5500ohm geophone is
a good indicator that something is wrong with the geophone itself. It still
responds ok to >0.1Hz, but <0.1Hz is just noise.



2012/8/2 Geoff 

> if the geophone is OK then i understand
> the rdamp should be like 4x Rgeo.
> together the final R value is (Rd||Rg)
> the final Ge is like GRg/(Rd + Rg).
> If your geophone has a resistance
> more than like 1% off at room temperature
> you must consider it damaged somehow.
> Regards,
> geoff
>
>
> On 2012-07-30 14:34, Tiago A. wrote:
>
>> Thanks for the answer Chris.
>>
>>     I am not sure that I understand what you have written - or what
>> system you are using. What ADC / number of bits, are you using ?
>>
>> The system (ADC) is a ADS1211P, the geophone is connected directly to
>> the ADC inputs with a 20cm wires. geIm using the A/D set to 22bits with
>>
>> the PGA set to 2x. With a Decimate Factor of 5600 with Turbo Mode 8,
>> gives a sample rate of 27.86, which sets -3dB at 8Hz (3rd order digital
>> filter) .
>>
>> You MUST damp a 1 Hz geophone correctly to get a flat output over 1 Hz.
>> What is the value of the damping resistor should i use? Its a 5500ohm
>> geophone. But the strangest thing is i can only measure 3800ohm (tested
>> today), i think when i got the geophone it had a higher resistance. This
>> geophone has be altered, the previous owner changed the original Bendix
>> 6/8 pin connector to a 4 pin Bendix connector.
>>
>> How are you getting your period expansion below 1 Hz down to 0.1 Hz ?
>> Only using AmaSeis. One of the few times i could get down to 0.03Hz was
>> in the Mexico M 7.4 earthquake (~7500km). It was offline in the Sumatra
>> earthquakes. I live in a small island in the middle of the ocean, so i
>> have always a lot of noise in the 2~5s period.
>>
>> What is the noise on your ADC - when you connect it to a resistor with
>> the same value as the geophone ?
>> I can check that in the next days.
>>
>>
>> Thanks.
>>
>> Regards,
>> Tiago.
>>
>>
>>
>> 2012/7/30 >
>>
>>     From: Tiago A. tiagoagre@......... 
>>
>>     Sent: Mon, 30 Jul 2012 3:08
>>     Subject: High noise >10s
>>     Hi.
>>
>>     Im using a L4 1Hz geophone and for almost a month im struggling with
>>     high noise from 10s up to ..., but from <10s it looks good. The
>>     noise comes and goes, this image is self explaining:
>>
>>     http://dl.dropbox.com/u/**15802004/noise.png
>>
>>     Changed cables, re centered the geophone, but didint help..the
>>     geophone is connected directly to a ADC without damping, external
>>     amplification or filtering, i only use digital filtering from the
>>     ADC. Before this issue start, i had a pretty clean spectrum >10s for
>>     almost 6 months, which allowed me to received LQ and LR from some
>>     strong events.
>>
>>     This sac contains 15minutes with high noise >10s, notice that at
>>     3:37 it recorded a Ml2.1 at 70km.
>>     http://dl.dropbox.com/u/**15802004/noise.sac
>>
>>     Hi Tiago,
>>          I am not sure that I understand what you have written - or what
>>     system you are using.
>>     You MUST damp a 1 Hz geophone correctly to get a flat output over 1
>> Hz.
>>     How are you getting your period expansion below 1 Hz down to 0.1 Hz ?
>>     Any period expansion DEPENDS on the geophone being correctly damped.
>>     What is the noise on your ADC - when you connect it to a resistor
>>     with the same value as the geophone ?
>>     You do need a low noise metal film resistor.
>>     What ADC / number of bits, are you using ?
>>     Regards,
>>     Chris Chapman
>>
>>
>>
>
> ______________________________**____________________________
>
> Public Seismic Network Mailing List (PSNLIST)
>
> To leave this list email PSNLIST-REQUEST@.............. with the body of
> the message (first line only): unsubscribe
> See http://www.seismicnet.com/**maillist.htmlfor more information.
>
Hi Geoff.

Increasing the sample rate to 125sps (-3dB at 36Hz), didin= t solve the issue and increased the >10s noise..i will try to increase m= ore the sample rate. But increasing the sample rate reveals a very high cou= nt at 50Hz, i check the earth connections it looks all ok..if i remove the = geophone from the ADC, i have +-2 counts of noise at 16bits and the 50Hz sp= ike is gone. Simply touching a wire from earth in any part of the geophone,= will make the geophone stiff, without responding.

SAC file with 15minutes at 125SPS
http://dl.dropbox.com/u/15802004/aa.sac

Nether= less the 50Hz issue, getting only 3800ohm from a 5500ohm geophone is a good= indicator that something is wrong with the geophone itself. It still respo= nds ok to >0.1Hz, but <0.1Hz is just noise.



2012/8/2 Geoff <= ;gmvoeth@.........>
if the geophone is OK then i understand
the rdamp should be like 4x Rgeo.
together the final R value is (Rd||Rg)
the final Ge is like GRg/(Rd + Rg).
If your geophone has a resistance
more than like 1% off at room temperature
you must consider it damaged somehow.
Regards,
geoff


On 2012-07-30 14:34, Tiago A. wrote:
Thanks for the answer Chris.

=A0 =A0 I am not sure that I understand what you have written - or what
system you are using. What ADC / number of bits, are you using ?

The system (ADC) is a ADS1211P, the geophone is connected directly to
the ADC inputs with a 20cm wires. geIm using the A/D set to 22bits with
the PGA set to 2x. With a Decimate Factor of 5600 with Turbo Mode 8,
gives a sample rate of 27.86, which sets -3dB at 8Hz (3rd order digital
filter) .

You MUST damp a 1 Hz geophone correctly to get a flat output over 1 Hz.
What is the value of the damping resistor should i use? Its a 5500ohm
geophone. But the strangest thing is i can only measure 3800ohm (tested
today), i think when i got the geophone it had a higher resistance. This geophone has be altered, the previous owner changed the original Bendix
6/8 pin connector to a 4 pin Bendix connector.

How are you getting your period expansion below 1 Hz down to 0.1 Hz ?
Only using AmaSeis. One of the few times i could get down to 0.03Hz was
in the Mexico M 7.4 earthquake (~7500km). It was offline in the Sumatra
earthquakes. I live in a small island in the middle of the ocean, so i
have always a lot of noise in the 2~5s period.

What is the noise on your ADC - when you connect it to a resistor with
the same value as the geophone ?
I can check that in the next days.


Thanks.

Regards,
Tiago.



2012/7/30 <
chris= atupw@....... <mailto:chrisatupw@.......>>

=A0 =A0 From: Tiago A. tiagoagre@......... <mailto:tiagoagre@.........>
=A0 =A0 Sent: Mon, 30 Jul 2012 3:08
=A0 =A0 Subject: High noise >10s
=A0 =A0 Hi.

=A0 =A0 Im using a L4 1Hz geophone and for almost a month im struggling wit= h
=A0 =A0 high noise from 10s up to ..., but from <10s it looks good. The<= br> =A0 =A0 noise comes and goes, this image is self explaining:

=A0 =A0 http://dl.dropbox.com/u/15802004/noise.png

=A0 =A0 Changed cables, re centered the geophone, but didint help..the
=A0 =A0 geophone is connected directly to a ADC without damping, external =A0 =A0 amplification or filtering, i only use digital filtering from the =A0 =A0 ADC. Before this issue start, i had a pretty clean spectrum >10s= for
=A0 =A0 almost 6 months, which allowed me to received LQ and LR from some =A0 =A0 strong events.

=A0 =A0 This sac contains 15minutes with high noise >10s, notice that at=
=A0 =A0 3:37 it recorded a Ml2.1 at 70km.
=A0 =A0 http://dl.dropbox.com/u/15802004/noise.sac

=A0 =A0 Hi Tiago,
=A0 =A0 =A0 =A0 =A0I am not sure that I understand what you have written - = or what
=A0 =A0 system you are using.
=A0 =A0 You MUST damp a 1 Hz geophone correctly to get a flat output over 1= Hz.
=A0 =A0 How are you getting your period expansion below 1 Hz down to 0.1 Hz= ?
=A0 =A0 Any period expansion DEPENDS on the geophone being correctly damped= ..
=A0 =A0 What is the noise on your ADC - when you connect it to a resistor =A0 =A0 with the same value as the geophone ?
=A0 =A0 You do need a low noise metal film resistor.
=A0 =A0 What ADC / number of bits, are you using ?
=A0 =A0 Regards,
=A0 =A0 Chris Chapman